Diagnostic Repair Actions - Philips FM-2 Service Manual

Antepartum fetal monitor
Table of Contents

Advertisement

Monitoring Degradation Problems

Diagnostic Repair Actions

Describe
Degraded
Performance
Swap
Transducers
Exchange the
Monitor
36
Examine the display after an apparently successful power up sequence and self-
test.
Consider, first, the appearance of artifacts, gaps, or general quality of the display.
Note the values being reported in the numerical frames (FHR(s)) and TOCO).
Are they clearly unreasonable?
Use a spare transducer that is free of defects and meets factory specifications. At
facilities where more than one monitor is installed, it may be convenient to
borrow a transducer that is working satisfactorily with another monitor.
If the monitoring problems persist, or there is total failure, exchange the
monitor.
Chapter 5 - Troubleshooting