Table 21
Test and Inspection Matrix
Test Block
Name
Visual Test
(See 20)
Power-On Self-
Test
(See 20)
Safety Tests:
•
Safety Test 1
(See 86)
•
Safety Test 2
(See 88)
Key to Table:
Chapter 8 - Test and Inspection Matrix
Test
Inspect exterior of fetal
monitor for damage.
Power on the monitor
Enclosure Leakage
Current - S.F.C. Open
Supply
Patient Leakage Cur-
rent - S.F.C. Open
Supply
P = Pass
Expected Results
No Visual Damage
Displays Normal Monitoring
Screen Configuration and
emits tones
All Safety Test results are in
range of Table 22 on page 87
All Safety Test results are in
range of Table 23 on page 89
F = Fail
When to Perform Tests Blocks
What to
Record
V:P or V:F
PO:P or PO:F
S:P or S:F
S:P or S:F
81