Attaching The Zero Calibration Box; Notes On Use Of Zero Calibration Box - Konica Minolta CM-3700A-U Instruction Manual

Table of Contents

Advertisement

Attaching the Zero Calibration Box

WARNING
Do not perform measurement with the specimen measuring port directed
towards your face. Doing so may cause damage to your eyes .
Be careful around openings in the CM-3700A-U. Failure to do so
may result in fingers being trapped causing injury.
The zero calibration box is used to perform zero calibration for measurement of reflectance.
{ Before performing zero calibration, attach the target mask (SAV) for calibration.

{ Before performing zero calibration, set the same specular component (SCI/SCE) as when performing

zero calibration using the software and ensure that UV light condition is set to "100% Full".
Procedure
1.
Pull the sample holder toward you and keep it
open.
{ The sample holder will remain open when

opened more than 70 degrees.
2.
Fit the projections of the zero calibration
box into the grooves on the CM-
3700A-U and then close the sample
holder to hold the box in place.

 Notes on Use of Zero Calibration Box


{ Take care not to scratch, touch, or make the inside of the zero calibration box dirty with such as fingerprints.

{ If the inside of the zero calibration box gets dirty, wipe it with a soft, clean, dry cloth.

{ If dirt is difficult to remove, dampen a cloth with commercially available lens cleaning liquid and wipe the

zero calibration box. Then wipe off the liquid with a cloth dampened with water, and leave the box to dry.
{ Should the inside of the zero calibration box get so dirty that it cannot be cleaned, replace the box with a

new one.
19