Operating Manual
pH electrodes
GE xxx
www.greisinger.de
Keep for later use.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GE Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GREISINGER GE Series

  • Page 1 Z03.0.0X.6C-11 English Operating Manual pH electrodes GE xxx Companies / Brands of the GHM www.greisinger.de Keep for later use.
  • Page 2: Table Of Contents

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 2 of 12 Index GENERAL NOTE ............................2 SAFETY ............................... 2 ............................2 NTENDED ..........................3 AFETY GUIDELINES ........................3 AFETY SIGNS AND SYMBOLS ..........................4 ORESEEABLE MISUSE ..........................4 AFETY INSTRUCTIONS BASICS ................................ 6 ..........................
  • Page 3: Safety Guidelines

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 3 of 12 The manufacturer will assume no liability or warranty in case of usage for other purpose than the intended one, ignoring this manual, operating by unqualified staff as well as unauthorized modifications to the device.
  • Page 4: Foreseeable Misuse

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 4 of 12 2.4 Foreseeable misuse The fault-free function and operational safety of the product can only be guaranteed if generally applicable safety precautions and the device-specific safety instructions for this document are observed.
  • Page 5 Z03.0.0X.6C-11 Operating manual GE xxx serie page 5 of 12 Do not use this product as safety or emergency stop devices or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage When interacting with chemicals at least the following points must be ensured: 1.
  • Page 6: Basics

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 6 of 12 3 Basics 3.1 pH measurement The pH value describes the acidic or alkaline behaviour of an aqueous solution. pH values below 7 are acidic (smaller values indicate higher acidity), and values above 7 are alkaline;...
  • Page 7: Further Information

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 7 of 12 The diaphragm, which establishes a connection between the electrolyte and the liquid to be measured, can be designed in different ways. Clogging or soiling of the diaphragm is a frequent cause of a malfunctioning or sluggish electrode.
  • Page 8: Calibration

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 8 of 12  For electrodes with liquid electrolytes: For continuous flow of electrolyte the cap sleeve made of rubber, that covers the refill opening, has to be removed before measurements. The opening has to be closed for storage in order to prevent running down. ...
  • Page 9: Care And Maintenance

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 9 of 12 4.3 Care and maintenance  For electrodes with liquid electrolyte: Check level of reference electrolyte and, if necessary, refill 3 mol/l KCl solution with help of an syringe or pipette. ...
  • Page 10: Choose The Right Electrode

    Z03.0.0X.6C-11 Operating manual GE xxx serie page 10 of 12 5 Choose the right electrode Sewage Aquarium water Soil testings Emulsions On-site measurements Fish farming Photo laboratory Galvanic baths Beverages Low-ion media (Rain water, some aquariums, VE water) Cosmetics Food sample Sea water Process chemistry Online measurements...
  • Page 11: Reshipment And Disposal

    We will then arrange for the proper and environmentally-friendly disposal. Private end users in Germany have the possibility of dropping off the product at the municipal collection centre. 7 Manufacturer / distributer GHM GROUP – Greisinger GHM Messtechnik GmbH Hans-Sachs-Str. 26 93128 Regenstauf...
  • Page 12: Specification

    Z03.0.0X.6C-10 Operating manual GE xxx serie page 12 of 12 8 Specification Reference Diaphragm type Type Description Operating range Pressure Connection Cable Notes Tube electrolyte /Membrane shape  GE 014-Cinch Cinch pH 2-12, 0..60°C 2x ceramic / low Cost pH-electrode pressureless 3mol/l KCl plastic approx.