H80.0.02.6C-16 Operating Manual GMH 5530 page 3 of 20 _____________________________________________________ _____________________________________________________________________________ 1 General Note Read this document carefully and get used to the operation of the device before you use it. Keep this document within easy reach near the device for consulting in case of doubt.
3. When connecting the device to other devices the connection has to be designed most thoroughly as internal connections in third-party devices (e.g. connection GND with protective earth) may lead to undesired voltage potentials that can lead to malfunctions or destroying of the GMH 5530 and the connected devices.
H80.0.02.6C-16 Operating Manual GMH 5530 page 5 of 20 _____________________________________________________ _____________________________________________________________________________ 4 Handling 4.1 Display elements Main display: pH value, ORP value (mV, mV rH value Secondary display: temperature value Arrows to selected measuring unit Rating of electrode state or battery status...
H80.0.02.6C-16 Operating Manual GMH 5530 page 7 of 20 _____________________________________________________ _____________________________________________________________________________ 5 Start Operation Connect electrodes, turn device on via key. After segment test the device displays some configuration: if zero point or slope correction is active (p.r.t. chapter “Configuration” and 9 “Input adjustment”) Remove protective cap from electrode.
First place pH and ORP electrode and temperature probe in the solution and stir carefully. 1. Measuring pH value: Connect the pH electrode and temperature probe to the GMH 5530. Then set device to pH measuring mode and calibrate electrode if necessary (p.r.t.
H80.0.02.6C-16 Operating Manual GMH 5530 page 9 of 20 _____________________________________________________ _____________________________________________________________________________ 6.4 pH electrode 6.4.1 Design In most cases so-called combination electrodes are used. That means that all needed elements are integrated in a single electrode (including reference electrode). Sometimes even a temperature sensor is integrated.
H80.0.02.6C-16 Operating Manual GMH 5530 page 10 of 20 _____________________________________________________ _____________________________________________________________________________ 6.5 Calibration of pH measurement The electrode data of pH electrodes are subject to fluctuation due to ageing and manufacturing tolerances. Therefore it is necessary to check the calibration with buffer solutions before measurements take place. If deviations are too large, a recalibration is necessary.
Page 11
H80.0.02.6C-16 Operating Manual GMH 5530 page 11 of 20 _____________________________________________________ _____________________________________________________________________________ 3. Calibration point 2: ‘Pt. 2’ (only for 2- or 3- point calibration) Place electrode and temperature probe No temperature sensor: (if any) in the second buffer solution (e.g.
H80.0.02.6C-16 Operating Manual GMH 5530 page 12 of 20 _____________________________________________________ _____________________________________________________________________________ 7 Configuration Some menu points depend on current device settings “menu” To change device settings, press for 2 seconds. This will activate the configuration menu (main display: “Set”). “menu”...
H80.0.02.6C-16 Operating Manual GMH 5530 page 13 of 20 _____________________________________________________ _____________________________________________________________________________ Menu Parameter Value Description Universal Output Interface and analog output off -> minimal power consumption SEr: Serial interface activated dAC: Analog output activated 01,11..91 Base address for serial interface communication...
H80.0.02.6C-16 Operating Manual GMH 5530 page 14 of 20 _____________________________________________________ _____________________________________________________________________________ The unit of all transmitter values (including measuring values) is the unit of corresponding displayed values. (e.g. temperature is displayed in °C -> temperature value is also transmitted in °C) 9 Input adjustment The zero point and slope of each measuring inputs can be adjusted with the parameters offset (“OFFS”) and...
H80.0.02.6C-16 Operating Manual GMH 5530 page 15 of 20 _____________________________________________________ _____________________________________________________________________________ 11 Real Time Clock (“CLOC”) The real time clock is used for chronological assignment of the calibration points. Please check the settings when necessary. 12 Accuracy Check / Adjustment Service You can send the device to the manufacturer for adjustment and inspection.
H80.0.02.6C-16 Operating Manual GMH 5530 page 16 of 20 _____________________________________________________ _____________________________________________________________________________ 14 Error and System Messages Display Description What to do? Battery empty Replace battery No display or Mains operation: wrong voltage or polarity Check power supply, replace it when necessary...
H80.0.02.6C-16 Operating Manual GMH 5530 page 19 of 20 _____________________________________________________ _____________________________________________________________________________ 17 Note A: temperature influence on pH buffer solutions GPH buffer capsules for 100 ml buffer solution Capsules for do-it-yourself mixing – unopened capsules can be stored a long time (approx. 3 years) T [°C]...