Figure 5 Install the shroud

7.4 Storage

The storage conditions for the LDO probe can be dry, wet, or humid. Wet storage conditions help to
keep the LDO sensor cap fully hydrated.
Note: After 72 hours of storage in tap water, the LDO sensor cap is fully hydrated.
For daily measurements in the laboratory, keep the probe in a humid environment such as a BOD
bottle filled approximately ¼ full with tap water. If the probe is used to monitor DO concentrations
continuously for more than 6 hours, keep the probe in tap water.
When the probe is used for short periods (less than 6 hours), the storage can be dry. Before dry
storage, rinse the probe in deionized water and blot dry. Soak the probe in tap water for 30 minutes
before use.
English 11