Control Systems Operator Test - Dräger Babytherm 8000 LDR Instructions For Use Manual

Warming bed
Table of Contents

Advertisement

Heat Therapy
Control
Systems
Operator Test
Control Panel
Checkout
20
°C
>37 °C
<35 °C
Figure 7. Operator Checkout of Control Systems
In order to check LEDs, displays and audible alarm:
Check
1. Press
button.
a. all LEDs illuminate for approximately 2 seconds
b. numeric displays show 88.8
c. an audible alarm is generated
d. displays and LEDs extinguish
e. the audible alarm is silent
f.
after 2 seconds, actual temperature is displayed.
This function check is possible during operation without affecting
Note:
any of the settings.
2. Perform control systems check at least once daily.
40 °C
Babytherm 8000
±1 °C
Inop
Check
!