Appearance Of The Irregular Heartbeat (Ihb) Symbol; Data Memory; Viewing The Stored Values; Memory Full - Microlife BR W3 Comfort Manual

Wrist blood pressure monitor
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
within the optimum (white), elevated (hatched gray) or high (black)
range. The classification corresponds to the following ranges defined by
international guidelines (ESH, ESC, JSH). Data in mmHg.
Range
Systolic Diastolic Recommendation
≥135
1. blood pressure
too high
2. blood pressure
130 - 134 80 - 84
elevated
3. blood pressure
<130
normal
The higher value is the one that determines the evaluation. Example:
a blood pressure value of 140/80 mmHg or a value of 130/90 mmHg
indicates «blood pressure too high».

Appearance of the Irregular Heartbeat (IHB) Symbol

This symbol AM indicates that an irregular heartbeat was detected. In
this case, the measured blood pressure may deviate from your actual
blood pressure values. It is recommended to repeat the measure-
ment.
Information for the doctor in case of repeated appearance of
the IHB symbol:
This device is an oscillometric blood pressure monitor that also
measures the pulse during blood pressure measurement and
indicates when the heart rate is irregular.
4
≥85
Seek medical
advice
Self-check
<80
Self-check

4. Data Memory

This device automatically stores the last 30 measurement values.

Viewing the stored values

Press the M-button 3 briefly, when the device is switched off. The
display first shows «M» AK and «A», which stands for the average of
all stored values.
Pressing the M-button again displays the previous value. Pressing the M-
button repeatedly enables you to move from one stored value to another.
Blood pressure readings with suboptimal cuff fit AO-A are not
considered in the average value.

Memory full

Pay attention that the maximum memory capacity of 30 memo-
ries is not exceeded. When the 30 memory is full, the oldest
value is automatically overwritten with the 31st value.
Values should be evaluated by a doctor before the memory
capacity is reached – otherwise data will be lost.

Clearing all values

If you are sure that you want to permanently remove all stored
values, hold down the M-button (the device must have been switched
off beforehand) until «CL ALL» appears and then release the button.
To permanently clear the memory, press the time button while «CL
ALL» is flashing. Individual values cannot be cleared.
Cancel deletion: press ON/OFF button 1 while «CL ALL» is
flashing.