General Information; Introduction - BeaconMedaes Medipoint 125 Operation And Maintenance Instructions

Medical gas central alarm
Table of Contents

Advertisement

Medical Gas Central Alarm - Medipoint 125 Digital Alarm
Safety Notice
Persons undertaking any installation and/or maintenance
must be fully trained in specialist work of this nature.
The "PERMIT TO WORK" procedure must be adhered to
for all installations once commissioned.
The alarm is designed and built in accordance with HTM
02-01 and ISO 7396-1 regulations and therefore should
be installed as such.
1.

General Information.

1.1

Introduction.

The BEACONMEDÆS Medipoint 125 digital (MP125D)
medical gas cental alarm is suitable for both the United
Kingdom and International markets, and fully satisfies
the requirements the requirements of the HTM02-01,
HTM2022 and ISO7396-1 applications. .
The Central Alarm is defined in ISO 7396-1 as ''Operating
Alarms'. The intention of the alarm is to indicate to
technical staff that it is necessary to replenish the gas
supply or to correct a malfunction within the Medical Gas
Pipeline System that requires an immediate response.
Each alarm panel requires a standard AC electrical power
supply of 115/230 volts ±10% at 50 or 60 Hz, and fused
at 3 Amps.
The central alarm can be programmed from on screen
menus to provide a fully flexible alarm system. Allowing
the panels to be adapted to suit any hospital's medical
gas monitoring requirements (see section 1.12 for
programming details).
An individual central alarm panel has 5 inputs or up to
Figure 1 - Medipoint 125 Digital Alarm (MP125D).
1
2
3
4
6 columns for displaying as a repeater panel consisting
of 'normal' and 4 'fault' conditions. These columns can
be setup to monitor either up to 5 medical gas source
equipment, or up to 20 individual point alarms, or a
combination of the 2.
One central alarm panel is designated the master panel
for control purposes. Any number of panels may be
programmed as a 'central acknowledge panel' and cause
all flashing displays on the system to change to steady
when the MUTE switch on the central acknowledge
panel is operated.
Each central alarm panel incorporates flashing alarm
displays, a POWER ON indication, a SYSTEM ALARM
indication, MUTE and TEST switches and an audible
warning programmed to operate depending upon the
Medipoint 125 consists of: -
No.
Description
Control Fascia
1
Power on LED.
2
Temperary mute button.
3
System Fault LED.
4
Test button.
LCD Touch Screen
5
Information screen.
6
Data logging screen.
7
Settup Menus.
8
Alarm service location.
9
Emergency battery supply status.
10
Gas service type
11
Gas service colour identification bar
12
Gas service status area
13
Indication that date/time is not set
14
Mute activated
13
14
5
5
6
7
4233500107.03
9
10
11
8
12