Exit Setup Screen - BeaconMedaes Medipoint 125 Operation And Maintenance Instructions

Medical gas central alarm
Table of Contents

Advertisement

4.4
Bar
NORMAL
Medical Gas Central Alarm - Medipoint 125 Digital Alarm
SCALE START:
-1.0
SCALE END:
16.0
3.5.8
Delete gas Badge.
ZERO
Click the "Bin"
icon to delete the gas badge
3.5.9
Confirm/reject changes.
Click the tick
to confirm and save the changes, which
will return you back to the main gas service setup screen.
The main screen will show a sample of the newly setup
gas service badge, and move the new badge icon to the
next location, see figure 33 example. Click the cross
to cancel the changes and return to the main gas service
setup screen.
Figure 33 - Main Gas Service Setup screen.
Special Characters/Icons:
Icon
Description
Previous Icon.
Takes you to the previous menus screen.
Next Icon.
Takes you to the next menus screen.
Exit Icon.
Takes you back to the main screen.
Add Gas service Badge.
Takes you back to the gas service setup screen.
Click on the New Gas Badge icon to start the process for
setting up the next gas service badge. Follow the steps
from section 3.5 until all the gas service have been setup,
up to a maximum of 5 inputs or a combination of inputs
and repeaters for a maximum of 6.
3.6
Exit Setup Screens
Once all the gas service badges have been setup press
on the "Exit"
icon to return to the main operating
screen. See figure 34 fro example of completed 6 gas
alarm.
INPUT:
1
CUSTOM:
LOW T/H:
3.6
HIGH T/H:
4.9
Figure 34 - Example, 6 gas main screen.
23