Additional Technical Functions - BeaconMedaes Medipoint 125 Operation And Maintenance Instructions

Medical gas central alarm
Table of Contents

Advertisement

Medical Gas Central Alarm - Medipoint 125 Digital Alarm
5.2

Additional Technical Functions.

The following section details instructions for technical
staff for additional diagnostics. Figure 41 details the
technical interface accessed from the touch screen.
Figure 41 - Example, 6 gas setup. Additional
technical functions.
Touch Screen - Functional Icons:
Icon
Description
Info
Press - Go to Info screen.
Press - Go to logs screen. Password
Logs
controlled (Factory default "1234".
Press - Go to setup menus. Password
Settings
controlled. See section 3.
5.2.1
Information Screen.
The information screen can be accessed to view the
device software version. See figure 42.
Press on the Info
icon from the main screen to access
the information screen.
Figure 42 - Information Screen.
Special Characters/Icons:
Icon
Description
Exit Icon.
Takes you back to the main screen.
The v#.#.# number indicates the software version.
Note - Figure 42 shows an example only, not the version
for the supplied device.
Press the exit
icon to return back to the main screen.
Note - The system will return back to the main screen
automatically if a new alarm condition becomes active,
or after one minute if the alarm is not manually exited
from the information screen.
5.2.2
Data Logging Screen.
The data logging screen can be accessed to view the
history of events for medical gas pipeline pressure faults
and user access to the settings menus. See figure 43.
Press on the logs
access the data logs. The first screen shows all entries.
Figure 43 - Data logging Screen, All data entries.
Special Characters/Icons:
Icon
Description
Previous Icon.
Takes you to the previous logs screen.
Next Icon.
Takes you to the next logs screen.
Down Icon.
Jumps down 50 records on the current logs screen
Up Icon.
Jumps up 50 page on the current logs screen
Scroll Bar.
Press, hold and drag the scroll bar navigate between
the selected 50 records.
Exit Icon.
Takes you back to the main screen.
Press on the next
through the alternative data logging screen. After the
main all entries data screen, the next set of screens
will be filtered by the medical gas service faults, in the
older of gas input from Column 1 through to Column 6
depending on the number of gasses setup on the alarm.
See figure 44 for an example of the filtered by gas fault
screen.
31
icon from the main screen to
or previous
icon to cycle
4233500107.03