Flowrate Calibration And Calibration Tables - Siemens SITRANS FUE1010 IP65 NEMA 4X Operating Instructions Manual

Ultrasonic flowmeters
Table of Contents

Advertisement

To Clear All Saved Data using the RS-232 Interface
NOTICE
Loss of RAM Data
Before proceeding further it is essential to understand that this function eliminates ALL data
stored in RAM. This means that all saved site setups including the site data of a flow-
calibrated site will be erased! In addition, the entire Datalogger file plus any custom factory
or user-created pipe or sensor tables will be eliminated.
The impact of this is such that we strongly recommend that you consult Technical Services
before continuing with this procedure. Be aware that you will have to create a new Site
Setup, re-enter all site specific parameters including pipe or sensor tables, plus all desired
Meter Facilities menu entries.
1. Turn off power (if it is currently on).
2. Turn the power on. As soon as you apply power, type the @ character three times.
– The prompt: [Clr Active Memory?] appears at the top of the screen. Press the <Down
3. Press the <Right Arrow> and then the <Down Arrow> to switch the option list to: [Clr
Saved Data? Yes].
4. Press <ENTER> to clear all Saved Site Data, Datalogger Data, user created Pipe Data
and Sensor Data.
5. To restore operation, press <MENU> to access the Installation Menu. Create a new site
setup or recall a stored site setup. Reselect any Meter Facilities items (e.g., RS-232 setup
parameters).
A.5

Flowrate Calibration and Calibration Tables

Flowrate Calibration Methods
SITRANS F 1010 equipment provides three ways to condition the calibration performance of
its flowrate output: Intrinsic (factory set), Kc, and Calibration Tables 1 through 3. Access to
these calibration options is found in the [Calibrate Flowrate] menu cell and the three [Calib.
Table] menu cells of the [Span/Set/Cal] menu.
Intrinsic
When selected, the flow meter uses no slope adjustment at all. Output data is still zeroed
and corrected for Reynolds number, but no slope adjustment is imposed on the flow meter's
flow register.
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
Arrow>.
Note
Note that the prompt switches to [Clr Saved Data? No].
A.5 Flowrate Calibration and Calibration Tables
Appendix
205