Connecting
5.5 Sensor Installation
Figure 5-21
1. Perform all required menu programming steps up until the point where you respond to the
[Install Complete?] prompt.
2. Make a note of the reported Number Index displayed in the [Install Sensor] menu. Check
to ensure that you have a matched set of sensors. They both should have the same serial
number but marked with either an "A" or "B" (e.g. 100A and 100B).
Note
Some sensors require a right-angle adapter. This adapter should be installed before
placing the sensors in the tracks.
74
To SITRANS F Transmitter
7ME39600CK Series Cable (992CNF)
Mounting Strap Guide
Sensor Clamp
7ME3950 Series Sensor Upstream
7ME39600M Series Mounting Track
(1012TN, 1012TNH)
Direct Mount 180° opposed with Mounting Tracks
7ME3950 Series Sensor Downstream
Flow direction
7ME39600M Series Mounting Track
(1012TN, 1012TNH)
REF Hole Index Pin
Ultrasonic Couplant
Pipe
Operating Instructions, 12/2014, A5E03086491-AC
FUE1010 IP65 NEMA 4X