Siemens SITRANS FUE1010 IP65 NEMA 4X Operating Instructions Manual page 203

Ultrasonic flowmeters
Table of Contents

Advertisement

Data Display Mode
After you complete the installation, you can toggle between Installation Menu mode to Data
Display mode. This is the same as using the <MENU> key on the keypad (see manual). The
PC keyboard equivalent to the <MENU> key is <CTRL> + <L>. Note that the RS-232
interface does not support graphics. Therefore, when you use HyperTerminal to view the
data display screens, you will see the same data in alphanumeric form only (as shown
below). You can still use the <Up Arrow> and <Down Arrow> to switch between available
display screens.
FUE1010 IP65 NEMA 4X
Operating Instructions, 12/2014, A5E03086491-AC
Current active site setup name. The [1]
indicates that the measurement channel is
active.
The current measured liquid sonic velocity.
A.4 RS-232 Connection
The flow Total display.
The current flow reading and flow units.
Appendix
201