Problem/possible cause
SmartStop is enabled, but does not automatically stop when I remove the mask.
Incompatible mask being used
Device Messages
Problem/possible cause
System fault, refer to user guide, Error 4
Device may have been left in a hot environment.
Air filter may be blocked.
There may be water in the air tubing.
All other error messages, for example, System fault, refer to user guide Error X
An unrecoverable error has occurred on the device.

General warnings

WARNING
Any change or modification to the product is not expressly approved by ResMed and could void
the user's authority to operate the device.
Supplemental oxygen must not be used while smoking or in the presence of an open flame.
When using the device with an oxygen supply, check the following:
Starting therapy – ensure the device is on and blowing air before the oxygen supply is turned
on.
Stopping therapy – ensure the oxygen supply is turned off first, then the device.
This will ensure oxygen does not accumulate within the device and create a risk of fire.
The device has not been tested or certified for use in the vicinity of X-ray, CT or MRI equipment.
Do not bring the device within 13 ft (4 m) of X-ray or CT equipment. Never bring the device into
an MR (Magnetic Resonance) environment.
The device should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the device should be observed to verify normal operation in the
configuration in which it will be used.
The use of accessories other than those specified for the device is not recommended. These may
increase radio frequency energy or be influenced by the interference and result in improper
operation.
For any serious incidents that occur in relation to this device, these should be reported to ResMed and the
competent authority in your country.
Solution
Only use equipment recommended by ResMed.
Contact ResMed or see ResMed.com for more information.
If you are using a nasal pillows mask with set pressure less than 7 cm
H
O (7 hPa). SmartStop will not work and should be disabled.
2
If you are using a conduit mask, SmartStop will not work and should be
disabled.
Solution
Allow to cool before re-use. Disconnect the AC adaptor and then
reconnect it to restart the device.
Check the air filter and replace it if there are any blockages.
Disconnect the AC adaptor and then reconnect it to restart the device.
Empty the water from the air tubing. Disconnect the AC adaptor and
then reconnect it to start the device.
Contact your care provider. Do not open the device.