ResMed AieSense 11 User Manual page 25

Table of Contents

Advertisement

Wireless module
Technology used
2G
3G
4G LTE Cat 1
*(bands may not be available in all regions)
FCC ID: 2ACHL-AIR114G
IC: 9103A-AIR114G
The AirSense 11 device complies with FCC Rules and Industry Canada rules. Operation is subject to the following two conditions:
1. This device may not cause harmful interference; and
2. This device must accept any interference received, including interference that may cause undesired operation.
The AirSense 11 device should be installed and operated with minimum distance of 0.59" (15 mm) between the equipment and the
user's body.
Additional information regarding the FCC Rules and IC compliance for this device can be found on ResMed.com/downloads/devices
In Canada, the device has been designed to comply with safety standards to radio waves (SAR) in accordance to RSS-102.
Temperature
Maximum heater plate temperature:
Temperature Cut-out (heater):
Maximum gas temperature: (at mask)
Recommended water type to use in the humidifier
tub (Standard tub)
time between each refill of the humidifier tub
1
The air flow for breathing produced by this therapy device can be higher than the temperature of the room. Under extreme
ambient temperature conditions (104ºF/40ºC) the device remains safe.
* tested at 23 ±2°C
General
The patient is an intended operator.
Humidifier performance:
SlimLine/Standard tubing
Mask Pressure
cm H
O (hPa)
2
4
10
20
1
Nominal % RH when
delivered to mask at
72ºF (22ºC)
Setting 4
80%
80%
80%
Frequencies (MHz)
900/ 1800 *
850/ 900/ 1700/ 1900/ 2100*
700/ 850/ 1700/ 1900*
154ºF (68ºC)
165ºF (74ºC)
≤ 106ºF (41ºC)
Distilled water (Americas only)
> 8 hours ± 0.5 hours*
Nominal % RH
delivered to mask,
tested in at 72ºF (22ºC)
ambient
Setting 8^ (maximum
setting)
100%
100%
100%
Max RF power output (dBm)
33.0
23.5
23.0
Nominal system output mg/L AH
Setting 4
≥6
≥6
≥6
1
2
, BTPS
Setting 8 (maximum
setting)
≥12
≥12
≥12
English
21

Advertisement

Table of Contents
loading

Table of Contents