Sound
Pressure level measured according to ISO 80601-2-70:2015 and ISO 80601-2-79:2018:
Device with SlimLine and humidifier tub (tub half filled) : 27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 and ISO 80601-2-79:2018:
Device with SlimLine and humidifier tub (tub half filled): 35 dBA with uncertainty of 2 dBA
Physical - device and humidifier tub
Dimensions (H x W x D):
Air outlet:
Wight (device and standard humidifier tub):
Weight (device and cleanable humidifier tub):
Housing construction:
Hot plate - material:
Water capacity:
Standard humidifier tub - material:
Cleanable humidifier tub - material:
Air filter
Standard:
Hypoallergenic:
Design life
Device, power supply unit:
Cleanable humidifier tub:
Standard humidifier tub
Air tubing:
Electromagnetic compatibility
The AirSense 11 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC 60601-1-2:2014,
for residential, commercial and light industry environments.
Portable and mobile RF communications equipment should be used no closer to any part of the machine, including cables, than the
recommended 3.94" (10 cm) separation distance.
The AirSense11 has been designed to meet EMC standards. However, should you suspect that the device performance (eg.
pressure or flow) is affected by other equipment, move the device away from the possible cause of interference.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found in
ResMed.com/downloads/devices.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Aircraft use
ResMed confirms that the machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M; RTCA-DO-160, section 20, category T) for all phases of air travel.
Maximum single fault steady state pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
40 cm H
O (40 hPa) for more than 1 second.
2
5.45" x 10.21" x 3.72"
(138.5 mm x 259.4 mm x 94.5 mm)
The 22 mm conical outlet connector complies with EN ISO 5356-1:2015
40 oz (1130 g)
40 oz (1130 g)
Flame retardant engineering thermoplastic
Stainless steel
380 mL
Injection molded plastic, stainless steel and silicone seal
Injection molded plastic, stainless steel and silicone seal
Material: Polyester non woven fiber
Average arrestance: >75%, when tested to EN779.
Material: Blended synthetic fibers in a polypropylene carrier
Efficiency: >80% (av) when tested to EN13274-7.
Note
: The use of a ResMed approved hypoallergenic filter will result in a
small reduction in the accuracy of the delivered pressure at high leaks.
5 years
2.5 years
6 months
6 months
English