6.2.1 Fixing unit for profile rods and wire-bar applicators

Put the wire-bar or profile rod into the holder (9).
Fix the fixing unit for profile rods and wire-bar applicators as shown below.

6.3 Application with gap applicator

When using the glass plate (i.e. in case of non-heatable applications) note the
followings:
Make sure that the glass plate is mounted with its glass surface facing up.
Alternatively a precision-vacuumplate can also be used.
Set the start stop (1) as follows:
Place your substrate (e.g. test chart / film) at the desired place on the glass
plate. Keep in mind that the gap applicator will be placed in front of the weight
(7).
Set your gap applicator to the desired gap height and place it on the substrate
at the desired starting position of the application area.
Lower the weight (7) to the bottom position.
Move the drawing unit (22) to the gap applicator.
Open the clamping device (5) and loosen the screw of the start stop (1).
Move the start stop (1) directly behind the weight (7) and refasten it with the
screw.
Fix the substrate with the clamping device (5), by pulling on the clamping grip
(3).
Hint: If the substrate is too small to use the clamping device (5), use a piece of
adhesive tape or the optional precision-vacuumplate to fix it.
Set the end stop (2) as follows
Loosen the screw of the end stop (2).
Move the drawing unit (22) forward and stop it with the end stop (2) as soon as
the desired drawing area has been reached.
: