Potentially dangerous situation Material damage Danger signs Meaning Source and possible consequences of non-observance Personal injury Physical injury, death Electrical voltage causes serious physical injury or Electrical voltage death Hot surface Hot surface can cause physical injury OME K04 04/17...
Procedures mechanical installation Procedures electrical installation Check or fault table Request for action Further associated documents Calibration certificate Declaration of conformity as per EU Directive 2006/42/EC Manufacturer's declaration as per EU Directive 97/23/EC Corresponding operating instructions for electronic equipment OME K04 04/17...
Proper use Safety □ Use volumeters of the OME series solely for flow measurement of lubricating liquids that are chemi- cally neutral and do not contain any gas or solids content. □ Volumeters require the operation with clean liquids. If coarse contamination, solid particles in the liquid or abrasive fine particles occur during operation, the volumeter has to be protected additionally by a correspondingly dimensioned operating filter in the pipe system, see "Protect the volumeter...
3900 3105 nenn n (Q Number of poles K1 K-factor K1 [P/l] 1214 17.73 Milliliters per pulse K1 [ml/P] 0.824 3.12 12.8 56.4 Pulse frequency f1 (Q [Hz] f1 (Q nenn f1 (Q Tab. 1 Maximum values OME K04 04/17...
Defective connection box or cabling due to exceeding of the maximum temperature. ► Do not heat the connection box and corresponding cables above the maximum temperature. ► Observe the values in the corresponding table, see Tab. 5, page 15. OME K04 04/17...
Load capacity Load capacity Load capacity OME 15 100000 50000 10000 5000 2000 1000 1000 h 5000 h 10000 h 30000 h 100000 50000 10000 5000 2000 1000 A Short-time operation B Continuous operation C Pressure loss D Flow The values apply for lubricating liquids at temperatures of up to 125 °C. Abrasive and aggressive liquids reduce the durability.
Load capacity Load capacity OME 20 100000 50000 10000 5000 2000 1000 1000 h 5000 h 10000 h 30000 h 100000 50000 10000 5000 2000 1000 10,5 13,5 A Short-time operation B Continuous operation C Pressure loss D Flow The values apply for lubricating liquids at temperatures of up to 125 °C. Abrasive and aggressive liquids reduce the durability.
Page 13
Load capacity Load capacity OME 25 100000 50000 5000 10000 2000 1000 1000 h 5000 h 10000 h 30000 h 100000 50000 10000 5000 2000 1000 A Short-time operation B Continuous operation C Pressure loss D Flow The values apply for lubricating liquids at temperatures of up to 125 °C. Abrasive and aggressive liquids reduce the durability.
Page 14
Load capacity Load capacity OME 40 100000 50000 10000 5000 2000 1000 1000 h 5000 h 10000 h 30000 h 100000 50000 1000 10000 5000 2000 17,5 52,5 87,5 122,5 157,5 A Short-time operation B Continuous operation C Pressure loss D Flow The values apply for lubricating liquids at temperatures of up to 125 °C.
Description Description Function description Fig. 1 Structure of the volumeter OME series 1 Connection 4 Ball bearing 2 Connection box 5 Measuring housing 3 Measuring screw small 6 Measuring screw large As screw meters volumeters belong to the group of rotating displacement meters. The fluid current makes the measuring unit rotate.
4. Close the connection with a blind flange. After about 6 months storage check the oil level in the volumeter and if necessary top up oil. OME K04 04/17...
► Before disposing of the volumeter neutralize the residues of the pumped liquid in the volumeter. 1. Disassemble the volumeter. 2. Clean residues of the pumped liquid from the individual parts. 3. Dispose of or recycle individual parts in accordance with the locally applicable regulations. OME K04 04/17...
► Do not take apart the volumeter. Installing the volumeter Volumeters of the OME series can be operated in any installation position. Notice: Both flow directions are possible. The preferred flow direction is indicated on the rating plate by means of a bright arrow, see Fig.
3. The screw-in length of the piping may not exceed the threaded length of the volumeter, since the flow cross-section is narrowed and internal components can be damaged. OME K04 04/17...
▪ Knowledge of the electrotechnical safety guidelines □ Standardly the OME sensors should only be connected with a BEM electronic unit, alterna- tively with an electronic unit released by the manufacturer. □ The connecting lines for the connection box are to be shielded and laid separately from the supply lines.
Tab. 1, page 23. The mesh width of the operating filter is relevant to the size of the volumeter. Volumeter size Mesh width max. OME 15 – 20 [mm] OME 25 – 40 [mm] 0.34...
1. Disassemble and clean the volumeter before recommissioning - or - ► Heat up the liquid during standstill by means of a heating system. 2. Switch on the system. Under the requirements mentioned above, the volumeter is ready for operation at any time. OME K04 04/17...
The interval at which recalibration is actually required depends strongly on the operating condi- tions of the device. The manufacturer provides the factory calibration as standard. If higher requirements are placed on the measuring instrument, an accredited calibration according to EN ISO/IEC 17025 is also possible. OME K04 04/17...
Mounting instructions OME 1. Remove the socket screws 915.1*, take off end cover 070.1* and o-ring 739.1, see Fig. 1, page 26. 2. Pull the screw set with the ball bearings and distance sleeve 039 out of the measuring housing 128, paying attention to the configuration for later assembly, see Fig.
6. Fasten the cable gland at the cover of the connection box. Terminal strip Designation Terminal strip Designation Com. t 10 – 30 V Com. t Sig. 1 Q Sig. t Sig. 2 Q Tab. 2 Connection table connection box OME K04 04/17...
► Increase the system pressure. ► Reduce the temperature. Pulsations too high ► Use another feed pump. ► Carry out changes to the system. ► Use the OMG series. Back pressure too low ► Increase the back pressure. OME K04 04/17...
Page 30
► Check whether the fluid flows in the system (pump in operation, slide valve opened, etc.). ► Check whether shut-off devices before and after the volumeter are opened. Volumeter switched to bypass ► Switch the volumeter to through-flow. Tab. 1 Fault table OME K04 04/17...
Spare parts Spare parts Appendix Maintenance kit OME 15 – 40 Notice: The maintenance kit contains only the numbered parts and is only supplied completely. 739.2 739.1 817.4 817.3 817.2 817.1 Maintenance kit OME 15 – 40 Fig. 1 Qty.
36.8 G 1" M 12 G 1 1/4" M 14 – – G 1 1/2" M 16 M 18 – – * Reduced tightening M 20 torque when screwing M 24 into aluminum Tab. 1 Tightening torques OME K04 04/17...
Contents of the EU Declaration of Conformity The volumeters described in these operating instructions are machinery in the sense of the Directive 2006/42/EC. The original of the EU Declaration of Conformity is enclosed with the machinery at delivery. OME K04 04/17...