  • Page 1 Operating Instruction Humidity/Temperature- Measuring Instrument Model: AFH-G...
  • Page 2: Table Of Contents

    9. Order Codes ............... 9 10. Maintence and Instruction for use ........9 11. Calibration ................ 10 12. Dimensions............... 11 13. Declaration of Conformance..........12 Manufactured and sold by: Kobold Messring GmbH Nordring 22-24 D-65719 Hofheim Tel.: +49(0)6192-2990 Fax: +49(0)6192-23398 E-Mail: info.de@kobold.com...
  • Page 3: Specific Application

    AFH-G 3. Specific Application The AFH-G is to be installed only in the specified applications. Every usage which exceeds the specifications is considered to be non-specified. Any dam- ages resulting thereform are not the responsibility of the manufacturer. The user assumes all risk for such usage.
  • Page 4 Thus, for one individual fibre with a diameter of 3 µm, a short saturation time (several seconds) can be measured. Empirical investigations show that bundled or woven fibres, as are used here in the Kobold sensor, give rise to a longer period prior to saturation. This is because the individual fibres impede each other during water absorption and/or water loss, and the ensuing humidity does not register until later.
  • Page 5: Instrument Inspection

    In case of damage, please inform your parcel service/ forwarding agent immediately, since they are responsible for damages during transit. Scope of delivery: · Measuring Unit Model “AFH-G” · Operating Instructions 6. Mechanical connection Unavoidable sources of interference should be kept at a good distance from the control systems.
  • Page 6 The installation position should be chosen so as to prevent condensed water from entering the housing. The preferred installation position is "sensor vertically pointing downwards" or "sensor horizontal". A grommet in the probe prevents penetration of water in the installation positions described above. Wall mounting set Seite 6 AFH-G 07/02...
  • Page 7: Electrical Connection

    AFH-G 7. Electrical connection 7.1. Passive sensors 7.2. Active sensors AFH-G 07/02 Seite 7...
  • Page 8: Technical Data

    Connection terminals: for conductor cross-sections 0.5 mm Cable connection: Cable gland PG 13.5 Electromagnetic compatibility: Noise immunity: EN 50 082-2 emitted interference: EN 50 081-2 Housing: Protection: IP 64 Weight: approximately 0.4 kg Seite 8 AFH-G 07/02...
  • Page 9: Order Codes

    (this applies to all humidity sensors with hygroscopic measuring elements). Such substances are resin aerosols, lacquer aerosols, smoke depos- its etc. The water-resistant property of the Kobold sensors allows for cleaning using water. Solvents cannot be used for this purpose. A light-duty detergent is recommended.
  • Page 10: Calibration

    AFH-G 11. Calibration Kobold sensors are adjusted correctly at a room temperature of 23ºC and 50%rh at a mean air pressure corresponding to 430m NN. If, however, a further adjust- ment is necessary, the following procedure should be adhered to : ·...
  • Page 11: Dimensions

    AFH-G 12. Dimensions AFH-G 07/02 Seite 11...
  • Page 12: Declaration Of Conformance

    AFH-G 13. Declaration of Conformance We, KOBOLD-Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Humidity/Temperature- Measuring Instrument Model: AFH-G to which this declaration relates is in conformity with the standards noted below: DIN EN 60730-2-1 1993...