Please read these operating instructions before unpacking and putting the unit into operation. Follow the instructions precisely as described herein. The instruction manuals on our website www.kobold.com are always for currently manufactured version of our products. Due to technical changes, the instruction manuals available online may not always correspond to the product version you have purchased.
4. Regulation Use Any use of the OMG which exceeds the manufacturers specifications, may invali- date its warranty. Therefore, any resulting damage is not the responsibility of the manufacturer. The user assumes all risk for such usage. 5. Operating Principle The measuring principle of the KOBOLD screw-volumeter is positive displacement.
Manometer connections: should be approachable thread: R 1/4") Storing: incorrect storing can cause corrosion even seizing Attention: When the volumeter or installations are damaged, medium could flow out. To prevent consequential damage an appropriate alarm device should be installed. OMG K04/0922 page 5...
9. Electrical Connection The dry sleeve is mounted, adjusted and tested with the pick up inside as a whole unit, by KOBOLD. The pick-up insert, including milled nut and spring, are screwed into the dry sleeve by the customer (see included data sheet).
Venting: The system must be free of air (affects the accuracy) 11. Dismounting and Mounting OMG 15 This work should be done by competent personnel only. This instruction should be used as a support for this activity. If spindle set,...
Page 8
OMG - 20 / 25 This work should be done by competent personnel only. This instruction should be used as support for this activity. If spindle set or measuring casing will be replaced, volumeter must be recalibrated! Dismounting: do not turn sensor sleeve ...
Page 9
OMG - 40 / 50 This work should be done by competent personnel only. This instruction is only used as support for this activity. If spindle set or measuring casing will be replaced, volumeter must be recalibrated! Dismounting: do not turn sensor sleeve ...
Page 10
Fig.1a Fig.1b Pay attention to correct position of spacer! (Fig.2) turning moment for the screw: OMG 40: large spindle 25 Nm small spindle 10 Nm OMG 50: large spindle 49 Nm small spindle 25 Nm press on rolling bearings ...
Page 11
change of bearings: remove second groove nut with locking plate remove circlips and spacers pull off rolling bearings remove o-rings Clean all parts carefully, do not scratch sealing surfaces! Fig.2 Fig.4b Fig.4a Fig.4 Fig.3 OMG K04/0922 page 11...
Page 12
Fig.1a Fig.1b press on rolling bearings turning moments for the screws: OMG-1H: large spindle 200 Nm small spindle 49 Nm OMG-1F: large spindle 200 Nm small spindle 86 Nm mount circlips and spacers on measuring spindles ...
Page 13
Screw in the dry sleeve carefully until it slightly grazes the rotating pole wheel (you can hear it through soft strikes), then turn back and tighten screw. Screw in pick up insert into dry sleeve. Proof signal with oscilloscope. If necessary correct distance. OMG K04/0922 page 13...
new pick up insert signal defective contacts check contacts external interferences install cables (and measuring transducer) carefully distance to pole wheel not check signal with oscilloscope - correct the correct distance page 14 OMG K04/0922...
13. Maintenance KOBOLD-volumeters are free of maintenance. At high bearing load it is useful to change the rolling bearings after a certain service life. When high accuracy is required, it is further advisable to calibrate the device peri- odically. 14. Technical Information...
Page 16
3 m cable Hall-effect 10…30 V -40..+150°C 420 bar arcap 3 m PTFE cable IP 67 magnetic 10…30 V -40..+250°C 420 bar arcap incl. transmitter/ IP 65 (0..+50°C) cable box with 1 m PTFE cable Connector Transmitter page 16 OMG K04/0922...
(Cd, Hg, Li or Pb) of the heavy metal that is decisive for the classification as containing pollutants: 1. „Cd" stands for cadmium 2. „Hg" stands for mercury 3. „Pb" stands for lead 4. „Li" stands for lithium Electrical and electronic equipment OMG K04/0922 page 21...
19. Manufacturer's declaration We, Kobold Messring GmbH, Hofheim-Ts., Federal Republic of Germany, hereby declare that the type OMG screw volumeters are intended to be fitted into a machine or assembled with other machinery to form a machine in accordance with Directive 2006/42/EC including its amendments, and that ...
20. EU Declaration of conformance We, KOBOLD Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: Screw volumeter Model: OMG-... to which this declaration relates is in conformity with the standards noted below: EN IEC 63000:2018 Technical documentation for the assessment of electrical...
Page 24
Electromagnetic compatibility (EMC) - Part 6- 4: Generic standards - Emission standard for industrial environments Also, the following EC guidelines are fulfilled: 2014/30/EU EMC Directive Hofheim, 09 Nov. 2022 H. Volz M. Wenzel General Manager Proxy Holder page 24 OMG K04/0922...