Kobold KAL Operating Instruction

Calorimetric flow monitoring device
Hide thumbs Also See for KAL:

Advertisement

Quick Links

Operating Instruction
for
Calorimetric Flow Monitoring Device
Model: KAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KAL and is the answer not in the manual?

Questions and answers

Summary of Contents for Kobold KAL

  • Page 1 Operating Instruction Calorimetric Flow Monitoring Device Model: KAL...
  • Page 2: Table Of Contents

    11. Maintenance ....................10 12. Fault Signalling ...................11 13. Technical Information..................11 14. Order Codes ....................12 15. Dimensions ....................13 16. Declaration of Conformance ...............14 Manufactured and sold by: Kobold Messring GmbH Nordring 22-24 D-65719 Hofheim Tel.: +49(0)6192-2990 Fax: +49(0)6192-23398 E-Mail: info.de@kobold.com Internet: www.kobold.com...
  • Page 3: Note

    By usage in machines, the measuring unit should be used only then when the machines fulfil the EWG-machine guide lines. PED 97/23/EG for KAL-11.. and KAL-12.. In acc. with Article 3 Paragraph (3), "Sound Engineering Practice", of the PED 97/23/EC no CE mark.
  • Page 4: Regulated Use

    4. Regulated Use The KAL is to be installed only in the specified applications. Every usage which exceeds the specifications is considered to be non-specified. Any damages resulting therefrom are not the responsibility of the manufacturer. The user assumes all risk for such usage. The application specifications include the installation, start-up and service requirements specified by the manufacturer.
  • Page 5: Operating Principle

    • Ensure that the permitted maximum operating pressure and temperatures for the device are not being exceeded. (See point 13 Technical Information) Installation: Mount the sensor in the pipe (KAL-1...)- in a sleeve welded to the pipe (KAL-2...). Fitting position: The function of the sensor is not dependent on their fitting position, provided that the pipe is completely filled with flow medium.
  • Page 6: Electrical Connection

    220 VAC, 110 VAC, 24 VAC +/- 10%, 50-60 Hz DC voltage supply 24 VDC + 15%, - 10% 8. Limiting Contacts Only on KAL-E1.../KAL-E2... 1x change - over relay for flow speed Only on KAL-E3... 1x change - over relay for flow speed...
  • Page 7: Alignment / Balancing

    Caution! There must be no air bubbles on the sensor. Flow speed must be at 0. 9.4. Commission electronic circuitry (switch on supply voltage) 9.5. Only on model KAL-E-3...Set temperature potentiometer (2) to left-hand stop. 9.6. Zero point alignment. Briefly activate the "SET"key (5) (using a suitable object e.g.
  • Page 8 It is possible at any time to reset the electronic circuitry to the full range • Flow potentiometer (1) at right-hand stop • (Only on KAL E3; Temperature potentiometer (2) to left hand- stop) • Press the SET key (5) and hold in position until the FLOW LED stops The device has now been reset into basic functioning mode.
  • Page 9 KAL 07/04 Page 9...
  • Page 10 Only KAL-E-3... with more than 10 m cable between sensor and electronic circuitry, as well as increased accuracy when measuring temperature. Temperature alignment • Caution! Disconnect from power supply. • Connect full length extention cable (>10 m) (terminals 6&8) •...
  • Page 11: Commissioning

    (1) Potentiometer (2) Potentiometer Setting of flow switch Setting of temperature Zero point alignment-turn to left-hand stop switch point (only KAL E3) -Extending measuring range -turn to right-hand stop (3) LED trend display for flow speed and switch point (4) LED temp is...
  • Page 12: Fault Signalling

    4 cm/s ... 200 cm/s Medium temperature: -20° C to 80 °C (Option: KAL-...HT-20...+120 °C max. pressure: 100 bar Caution! KAL 1132 and 1140 max. 25 bar Protection (EN 60529): IP 68 Electronic system Power draw: max. 3,6 Watts Breaking capacity: max.
  • Page 13: Order Codes

    Flow LED will flash. Power failure: Calibration data retained for 10 years No Battery Trend display approx. 10% Consistency: ca. 1% Reaction time: generally 5 sec., max 12 sec. Accuracy of temperature limit: ± 2 °C 14. Order Codes KAL 07/04 Page 13...
  • Page 14: Dimensions

    SW 19 G 3/8 SW 27 G 1/2 SW 27 G 1/2 SW 27 G 3/4 SW 32 G 3/4 SW 32 82,5 SW 39 G 1 1/4 SW 46 G 1 1/2 SW 55 92,5 Page 14 KAL 07/04...
  • Page 15: Declaration Of Conformance

    16. Declaration of Conformance We, KOBOLD-Messring GmbH, Hofheim-Ts, Germany, declare under our sole responsibility that the product: KAL-E1... KAL-E3 Calorimetric flow monitoring device to which this declaration relates is in conformity with the standards noted below: EN 50081-1 EMC General Emission Requirements...