A device's ID is the serial number of the gauge, and the nominal width of the
gauge. Sometimes the device appears in the device list with the incorrect ID due
to a corruption in the register storing the device's ID. The gauge's ID can be
restored.
1. Make sure that the gauge is paired and connected to the computer/tablet.
2. Put the nominal gauge in mm and the gauge ID (serial number).
3. Select 'Left or only device'.
4. Press the 'Restore ID' button.
5. Press 'Unpair selected device' to unpair the Routescan
6. Restart both the computer/tablet and the Routescan. The Routescan will
now appear with its restored ID in the list of devices found.
9.10 Save Calibration DATA
If you wish save the calibration data from the gauge, press 'Save calibration'
button. It will be stored in C:\abtus\system\Calibration Files with the current date.
9.11 Disconnected warning message
This message means that the control unit has lost the Bluetooth connection with
the Routescan. It can be caused by:
Low Voltage Battery. You will see the battery led indicator blinking in red
in that case or you won't able to switch it on again because the battery is
totally discharged.