Keysight N9030B Installation Note page 7

Pxa signal analyzer, option yav y-axis video out
Hide thumbs Also See for N9030B:

Advertisement

Manually Verify Operation
1. Preset X-Series analyzer. Assure mode is Spectrum Analyzer.
2. Display the 50 MHz calibrator signal. Press Input/Output, RF Calibrator, 50 MHz. Press FREQ,
Center Frequency, key in 50 MHz. Press Span, Zero Span. You should see a flat line across the
display at −25 dBm.
3. Press AMPTD, Ref Level, and key in −25 dBm. The displayed signal level should be at the top
graticule line.
4. Press Input/Output, Output, Analog Out, Screen Video.
5. Connect a voltmeter, set to measure DC volts, to the analyzer rear panel ANALOG OUT
connector.
6. The voltmeter should measure 1V.
7. Increase the Ref Level to −15 dBm. The signal should drop one graticule division on screen and
the voltmeter reading should be 0.9 V.
8. This test is sufficient to determine if the output circuitry is working properly for all Analog Out
signal selections.
Installation Note N9030-90086
Option YAV Installation Kit