Atmos S 201 Thorax Short Operating Instructions page 2

Hide thumbs Also See for S 201 Thorax:

Advertisement

Warnings
Display
Filling the water lock
Key lock
After switching on the device the key lock is automatically activated if no key is operated within
the preselected period (factory setting: 5 min).
For manual activation of the key lock, please press keys
For manual activation of the key lock, please press keys
symbol is displayed on the screen.
For deactivation, please press the buttons
For deactivation, please press the buttons
is no longer displayed on the screen.
Grafi c display
In the grafi c display the course of therapy can be represented graphically.
In the grafi c display the course of therapy can be represented graphically.
With the
With the
key you can switch over between Flow, Time and Vac.
key you can switch over between Flow, Time and Vac.
The activated dimension is coloured.
With the keys
With the keys
, settings can be
, settings can be adjust
Flow scale: scalable between 0-1 & 0-20 l/min, start setting 0-1 l/min
Time scale: scalable between 0-30 min and 0-12 Tagen in 9 steps,
start setting 0-30 min
Vacuum scale: scalable between 0-10 & 0-100 mbar,
start setting 0-100 mbar
Cause of defect
If the fi nal vacuum cannot be reached an
acoustic and optical alarm is released and
„vacuum low" is displayed. Possible cause for this
error message: Leakage, blocked bacterial fi lter,
blockage in the measuring channel or secretion
in the pump.
When an excessive vacuum is measured the
„vaccuum high" alarm is activated (acoustic and
optical).
In case the battery voltage falls below a certain
value the alarm signal (acoustic and optical) for
„battery low" is activated.
The information that the system is blocked is
activated.
During normal operation the fi lling of the water lock is optional.
To fi ll the water lock a sterile cannula 20 G, a sterile syringe and 50 ml sterile
water are required. Take the cannula and puncture the silicone seal above the riser and
then fi ll the water lock until the 2 cm fi lling mark is reached.
The canister may only be removed when the pump is switched off .
simultaneously until the
simultaneously until the
adjusted
Check connections and hoses.
Check the hose system.
Connect the device to the mains,
battery is recharged and charging
condition is displayed.
Check the hose and in case the fi lter
is blocked, exchange the canister
and the fi lter.
simultaneously until the
simultaneously until the
symbol
symbol
Troubleshooting