Vibration - Agilent Technologies Varian 320-MS Pre-Installation Instructions

Quadrupole mass spectrometer lc/ms
Hide thumbs Also See for Varian 320-MS:
Table of Contents

Advertisement

Table 5 Additional Bench Space Requirements
Function
Provide space for air circulation, gas
lines, and electrical connections.
Dissipate room heat and allow for
routine maintenance.
Analytical Systems only:
Provide space for solvent bottles.
Please refer to the Varian LC Systems Pre-installation Instructions, part number 391483100, for
information about LC modules.
Use Table 6 and Table 7 to determine the required width of the bench and the weight the bench
must support for your configuration.
Regardless of configuration, the depth of the bench depends on the clearance between the
bench and the wall to accommodate the vacuum hoses for the foreline pumps.
Table 6 Bench Width Worksheet
Bench
Width
cm
in.
Table 7 Bench Load Worksheet
Bench
Load
kg
lb

Vibration

Ensure that the benches are free from vibrations, especially those caused by equipment in
adjoining locations. Because the foreline pump vibrates during operation, it belongs on the floor
below the LC/MS, not a on the bench with the LC/MS.
9300025700: Rev. A
If the distance between the bench and the wall is greater than or equal to 10.2 cm
(4 in.) then the bench must be at least 84 cm (33 in.) deep. The vacuum hoses will have
sufficient room to connect the MS to the foreline pump.
If the distance between the bench and the wall is less than 10.2 (4 in.) then the bench
must be at least 94 cm (37 in.) deep. To accommodate the vacuum hoses, two 5.8 cm
(2.3 in.) holes must be drilled into the bench on the left side and to the rear of the MS
location.
Auto-
___-LC
sampler
______
______
______
______
Check the checklist box: Bench space is sufficient.
Auto-
___-LC
sampler
Components
______
_______
______
______
Check the checklist box: Bench can support system weight.
320-MS with 460-LC, at least 114 kg (250 lb)
320-MS with HTS PAL, at least 103 kg (231 lb)
Check the checklist box: Bench does not vibrate.
Space Allowance
15 to 30 cm (6 to 12 in.) behind the system.
At least 76 cm (30 in.) above the system.
At least 30 cm (12 in.) to side of the LC.
Other
MS
Components
______
53.5
______
21
Other
MS
______
61
______
134
6 of 14
Monitor and
CPU
Keyboard
18.7
51
7.5
20
Monitor and
CPU
Keyboard
11.5
5.5
25.3
11.5
Pre-installation Instructions
Total
___cm
____in.
Total
____kg
____lb