Agilent Technologies Varian 320-MS Pre-Installation Instructions page 9

Quadrupole mass spectrometer lc/ms
Hide thumbs Also See for Varian 320-MS:
Table of Contents

Advertisement

LC/MS
The LC/MS requires the following.
Table 9 LC/MS Circuits, Power Plug and Outlet Types
Component
320-MS
MS40+ Foreline
Pump
Computer
LC modules
Table 10 LC/MS Power Requirements
Module
320-MS
MS40+ Foreline Pump
Use a separate dedicated power source for HPLC modules and additional instruments and
equipment. Never plug the mass spectrometer and the chromatograph into the same power
source or the power source may overload. Never use the free outlet on any of the power
sources for equipment that draws more than 2A.
Table 11 MS and Foreline Pump Power Cords
Power Cord
Length
Plug
The power cables for the CPU, monitor, and printer are approximately 2.13 m (7 ft) long.
Figure 3 shows the power plugs and outlet types.
9300025700: Rev. A
Circuit
1 each 220V
1 each 220V
110V or 220V, see appropriate user manual
110V or 220V see appropriate user manual
Max Current Draw (Amps)
1 each 2.5 m (8 ft)
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
Replacing or substituting power cords or plugs must be done
with strict compliance with all regulations, including electrical
CAUTION
codes, power cord color coding, and appropriate regulatory
agency certification marks.
Power Plug and Outlet Type
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
10A
10A
320-MS
9 of 14
MS40+ Foreline Pump
1 each 2.5 m (8 ft)
US: 1 each NEMA 6-15P
EU: 1 each CEE 7/7 plugs
Pre-installation Instructions