OTHER DISPLAY CODES
-PO- = Pump Off
-OC- = Open Circuit on temperature input
9.
FOR BEST PRESSURE SENSOR ACCURACY
1)
Switch the analyser on for 5 minutes to let the temperature stabilise.
2)
Zero the pressure sensor when the analyser in the exact position that it will be
used.
10.
MANO-MODE
A limited mode for pressure and temperature only that avoids the 90 second countdown
.
With the KANE457 switched off, press and hold down
and release
top line.
The KANE457 will now operate as a fixed display thermometer/pressure meter with the
pump off and inhibited. The readings can be printed but not stored.
11.
TO FULLY CHARGE NiMH RECHARGEABLE BATTERIES
1)
The analyser must be switched on.
2)
Connect the charger and switched it on; charging indicator illuminates.
3)
Switch the analyser off; the display will show "BATTERY CHARGING".
/
. Release
/
/
after MANO_MOD is displayed on
and then press