Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Let's now use the scope's Segmented Memory mode of acquisition to
selectively capture 500 consecutive occurrences of our trigger condition
(Read from address 29
must be licensed with the Segmented Memory option.
31
32
33
34
35
36
37
38
Figure 54 Using the scope's Segmented Memory acquisition to selectively capture more
Segmented Memory optimizes oscilloscope acquisition memory by only
capturing important segments of a waveform based on the trigger
condition and timebase setting. In this example, we have selectively
3000 X-Series Oscilloscopes Advanced Training Guide
HEX
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
Set the scope's timebase to 100.0 µs/div.
Press the [Run/Stop] front panel key to begin repetitive acquisitions
again.
Press the [Acquire] front panel key.
Press the Segmented softkey; then turn the Entry knob to select 500 as
the number of segments to capture.
Press the Segmented softkey to begin a Segmented Memory acquisition.
Press the Current Segment softkey; then turn the Entry knob to review all
500 captured segments.
Set the Current Segment to 500 and note the time- tag of the last captured
segment.
2
I
C traffic.
). But to complete the rest of this lab your scope
4