3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
occasional violations. Uniquely triggering on edge speed violation signals
can be accomplished on Agilent's 3000 X- Series oscilloscopes using the
Rise/Fall Time triggering mode.
11
12
13
14
15
16
Figure 33 Using Rise/Fall Time triggering to trigger on violation edge speeds.
Your scope should now be triggering on just rising edges of the input
signal that have rise times in excess of 100 ns as shown in
addition to triggering on Rise/Fall Time violation edges, the Agilent
52
Press the [Trigger] front panel key; then turn the Entry knob to select
the Rise/Fall Time trigger type.
Press the Level Select softkey to select Low.
Turn the Trigger Level knob to set the lower trigger threshold level to the
approximate 10% level. Note that you can visually use the lower
measurement threshold marker as a good indication as to where the
approximate 10% level is located.
Press the Level Select softkey until it indicates High; then turn the Trigger
Level knob to set the upper (high) trigger threshold level to the
approximate 90% level. Note that you can visually use the upper
measurement threshold marker as a good indication as to where the
approximate 90% level is located.
Press the "< >" softkey to select ">".
Press the Time softkey; then turn the Entry knob to set the violation
time to > 100 ns.
3000 X-Series Oscilloscopes Advanced Training Guide
Figure
33. In