4
Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
decoded record can be difficult. So let's now perform an automatic search
to find all occurrences of errors. We will then automatically navigate to
each of these error conditions.
22
23
24
25
26
Figure 76 Using the scope's automatic Search & Navigation capability to find all
The white triangles near the top of the waveform area mark the time
location of each "found" occurrence of our search operation as shown in
Figure
column of the lister table. Your scope should have found and marked more
than 15 occurrences of this search operation based on a total acquisition
time of 2 seconds. Note that the number of "found" events is indicated
near the bottom of the display. To automatically navigate to each LIN
error, press the
102
Push the horizontal position/delay knob to re- position the trigger point
back to center- screen.
Set the scope's timebase to 200.0 ms/div.
Press the [Search] front panel key.
Press the Search Edge softkey; then select Serial 1 (LIN).
Press the Search for softkey; then select Errors using the Entry knob.
occurrences of LIN errors over a 2 second time span.
76. These frames are also marked in orange in the first/pointer
and
front panel navigation keys.
3000 X-Series Oscilloscopes Advanced Training Guide