Setting The Power-On Initialization Values; Table 4-2: Factory Default Power On Settings - Ametek CS Series User Manual

Table of Contents

Advertisement

User Manual – Rev J
1. Press the Menu key repeatedly until the REGISTERS / CONFIGURATION menu is
displayed.
2. Move the cursor to the REGISTERS entry and press the ENTER key.
3. The cursor will default to the SAVE REGISTER # position. Enter a number from 1 through
15 and press the ENTER key. DO NOT USE REGISTER 0 (REG0) as it is reserved for
power-on setting recall and does not include a transient list.
4. A message will appear indicating that the front panel settings and the transient list data have
been saved in the setup register you selected.
4.8

Setting the Power-on Initialization Values

The power source is shipped with default factory settings when the unit is powered up. The
factory settings are:
Parameter
Current
Frequency
Compliance Voltage
Output state
Local / Remote State
It is possible to change the power on initialization values in one of two ways:
1. Using the IEEE-488, RS232C, LAN or USB bus interface.
2. Using the front panel.
To change the power on initialization values from the front panel, proceed as follows:
1. Set the AC power source output parameters from the front panel as you want to power
up the unit.
2. Save this setting to setup register 0 from the REGISTERS menu.
3. Select the CONFIGURATION menu and move to the POWER ON field.
4. Change the POWER ON field to REG0.
5. This will recall the settings contained in register 0 at power up.
California Instruments
Factory default setting
0.00 A
60 Hz
135 V
OFF
Local

Table 4-2: Factory Default Power on Settings