3.5.3
Multi-chassis Output Wiring Diagram
Figure 3-2 shows the required output connections for a 9000CS/2 two chassis system (rear-view
perspective). Always turn off AC mains power to the 9000CS/2 by turning off the circuit
breakers on both the Master and Auxiliary 4500CS power source before making or
changing output connections. The terminal block shown to connect the outputs of both
chassis together is provided in the 9000CS/2 ship kit. The System Interface cable is a DB25 to
DB25 M/F cable approximately 2 meters in length. (CI P/N 250778). This cable connects
between the male DB25 connector on the Master unit rear panel labeled TO AUXILIARY
INTERFACE and the female DB25 connector on the Auxiliary unit rear panel labeled TO
MASTER INTERFACE as shown in Figure 3-2. The OUTPUT SAFETY COVER must be
removed to use the System Interface and the AC Source must be installed in a cabinet with a
protective rear screen or door.
AUX OUTPUT
HI
LO
HI
LO
LOAD
øA
øB
øC
N
9000CS/2
TERMINAL
BLOCK
AUX OUTPUT
HI
LO
HI
LO
California Instruments
RS232C
IEEE-488
TRIGGER
FLT
INH
FSTB IN1
RPV
USB
LAN
OUTPUT TB1
ØA
26VAC
ØB
TO MASTER INTERFACE
SENSE
ØC
5VAC
ØA
ØB
TO AUXILIARY INTERFACE
N
ØC
N
300 VAC MAX TO
CLOCK
LOCK
RS232C
TRIGGER
FLT
INH
FSTB IN1
RPV
USB
LAN
OUTPUT TB1
ØA
26VAC
ØB
TO MASTER INTERFACE
SENSE
ØC
5VAC
ØA
ØB
TO AUXILIARY INTERFACE
N
ØC
N
300 VAC MAX TO
CLOCK
LOCK

Figure 3-2: 9000CS/2 Output Wiring

MASTER
SERIAL TAG
ØA
VOLTAGE INPUT
ØB
RATING
ØC
INPUT
SAFETY
COVER
AUXILIARY
SERIAL TAG
ØA
VOLTAGE INPUT
ØB
RATING
ØC
INPUT
SAFETY
COVER
CS Series
INPUT TB3
INPUT TB3
A
B
C GND
AC SERVICE
33

Advertisement

Table of Contents
loading

This manual is also suitable for:

3000cs4500cs9000cs/213500cs/318000cs