Output Programming - Ametek CS Series User Manual

Table of Contents

Advertisement

User Manual – Rev J
ENTRY
PHASE (C)
4.3

Output Programming

4.3.1
Set the Output
Output parameters are all set from the PROGRAM screen.
1. Use the MENU key and select the PROGRAM entry.
2. Press the ENTER key to bring up the PROGRAM menu.
or
2. Use the PROG key to directly bring up the PROGRAM menu.
There are two methods for programming output parameters:
IMMEDIATE mode
SET mode
4.3.2
Slewing Output Values with the Knob in IMMEDIATE Mode
The default mode of operation is an immediate mode in which changes to output parameters
made with the knob or the entry keypad are immediately reflected at the output.
To change the output current:
1. Place the cursor on the CURR entry
2. Rotate the shuttle knob clockwise to increase the value, counterclockwise to decrease the
value or use the Keypad to enter a value and press the Enter key.
These changes take effect immediately.
California Instruments
DESCRIPTION
In three-phase mode, available current is three
times limit shown.
For multi-chassis configuration, limit field still
shows current per chassis and phase.
Phase Setting
Displays phase angle for phase C. Valid values
are 120 for three-phase or mode configuration,
0 for single-phase only configuration. Any other
value indicates split (2) phase configuration.
Counter
Clock
Clock
wise
wise
DECR
INCR
CS Series
77