Disinfection

High Level disinfection
In the following procedures, only one disinfection process needs to be performed: Thermal disinfection OR
Chemical disinfection.
High level thermal disinfection
1. Immerse the components in a water
bubbles are trapped.
2. Soak the components in a hot water bath. ResMed has validated:
Water bath: at 167ºF (75ºC) for 30 minutes
Note:
Higher temperatures may damage the components.
3. Allow the components to dry out of direct sunlight.
OR
High level chemical disinfection
1. Make a solution of Ortho-phthalaldehyde 0.55% as directed by the disinfectant manufacturer. ResMed
has validated:
CIDEX® OPA Ortho-phthalaldehyde
2. Soak the components in the solution at room temperature (approximately 69.8ºF to 77ºF (21ºC to 25ºC)
for 12 minutes. Agitate the components in the disinfection solution to ensure there are no air bubbles.
3. Rinse and agitate the components in water
≤ 140ºF (≤60ºC) for 1 minute. Shake air tubing to remove excess water.
Repeat the rinse procedure two additional times using fresh water
1. Allow the components to dry out of direct sunlight.
Inspection
Perform a visual inspection of each component.If any visible deterioration is apparent (holes, tears or
cracks etc) replace the component.

Reassembling

Once the components are dry, reassemble the device.
Outlet connector
1. Hold the outlet connector with the seal pointing to the left and the clip pointing forward.
2. Make sure the outlet connector is correctly aligned and insert the outlet connector into the socket. It
will click in place.
For all cleaning, rinsing and disinfection steps use drinking quality water.
2
30
bath. Agitate the components in the water bath to ensure no air
2
2
5 liters per component at
2
for a total of three rinses.