Preparing the device for use between patients
When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health
care provider, the outlet connector, air tubing , cleanable humidifier tub and device enclosure should be
reprocessed between each patient.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the
outlet connector, air tubing, cleanable humidifier tub and device enclosure. The person executing
reprocessing activities is responsible for ensuring that reprocessing is completed in line with ResMed's
validated procedures. Components not identified in the reprocessing instructions do not require
reprocessing or are intended for single patient use.
Note: The standard HumidAir 11 tub cannot be reprocessed.
WARNING
• Always follow cleaning and disinfection instructions. Some cleaning products may damage
the components and their function or leave harmful residual vapors.
• Any deviations from the procedures or claimed maximum number of cycles in this guide can
have an adverse effect on the components and consequently the safety or the quality of
therapy.
• When using detergents, disinfectants or equipment, always follow the instructions provided
by the manufacturer of those products. In the event of conflict, this guide takes precedence.
• Always follow safe operating practices, including the use of appropriate Personal Protective
Equipment (PPE), as required. Refer to the instructions provided by the manufacturer of those
products for more details.
• Beware of electrocution:
• Do not immerse the device, AC Adaptor or power cord in water.
• Do not connect to power while the device is wet. Make sure that all parts are dry before plugging
it in.
1
Disinfection Guide

Advertisement

loading

Summary of Contents for ResMed AirSense 11

  • Page 1 Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the outlet connector, air tubing, cleanable humidifier tub and device enclosure. The person executing reprocessing activities is responsible for ensuring that reprocessing is completed in line with ResMed's validated procedures. Components not identified in the reprocessing instructions do not require reprocessing or are intended for single patient use.
  • Page 2 General summary ResMed has validated the following number of cycles for cleaning and disinfection using a manual cleaning method. Components Cleaning - Mild alkaline, anionic detergent (eg, Alconox™) System component Chemical high level disinfection Thermal high level disinfection eg, CIDEX ®...
  • Page 3 Air tubing 1. Pinch the cuff of the air tubing, and gently pull it away from the device. 2. Hold both the cuff of the air tubing and the swivel of the mask, then gently pull apart. Outlet connector 1. Locate the outlet connector on the inside of the device and release it by pressing the clip firmly. 2.
  • Page 4: Device Enclosure

    Device enclosure Cleaning Clean the device enclosure using an alcohol based cleaning and disinfection wipe. ResMed has validated CaviWipes1. 1. Wipe the exterior of the device using a wipe until visually clean following the manufacturer's instruction for cleaning. Use a minimum of two wipes.
  • Page 5 Agitate the components in the water bath to ensure no air bubbles are trapped. 2. Soak the components in a hot water bath. ResMed has validated: • Water bath: at 167ºF (75ºC) for 30 minutes Note: Higher temperatures may damage the components.
  • Page 6: Air Tubing

    1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia See ResMed.com for other ResMed locations worldwide. AirSense is a trademark and/or registered trademark of the ResMed family of companies. For patent and other intellectual property information, see ResMed.com/ip. Alconox is a trademark of Alconox INC.