2.2.6. Print Options

If the instrument includes a Printer, press Menu > Reports from the Main
screen. This screen allows the user to print the results of the last test, the
contents of stored methods, or the current instrument settings. The printer
is also enabled or disabled from this screen by using the Enable Printer
check box.
To print the results of the most recently executed method, ensure there is
a check in the Enable Printer box and press Last Run Results.
To print the parameters of any of the stored methods, ensure there is a
check in the Enable Printer box and press Print Method. Select the
number that corresponds to the method to be printed. Available methods
have a name associated with them. Methods yet to be configured have a
series of dashes (------) in the name field.
To print the current 708-DS instrument settings, ensure there is a check
in the Enable Printer box and press Print Instrument Settings. All the
instrument configuration parameters and their current values are
documented on the printout.

3. Measurement of Physical Parameters

3.1. RPM, Wobble, Centering, Level, Verticality (Shaft and Vessel)

Periodic verification of certain physical parameters is critical for all dissolution
apparatus. Each laboratory has specific guidelines and procedures to follow that
detail the frequency of these measurements. The following sections will suggest
how to perform these steps with the 708-DS dissolution apparatus. The primary
tools used in these descriptions will be Agilent's QA II Station and/or VK 5010.
Other similar calibrated devices may also be utilized.
Revision A
P/N: xx-xxxx
Agilent 708-DS Dissolution Apparatus
Figure 2-16. 708-DS Printer
- 18 -
Training Manual
September 2010