Aspect ISR G3 Q-CARD
WARNING! Read operator
SETUP/PATIENT PREPARATION:
Moving table:
To move table, set caster lock pedal to FLOAT or
STEER position.
To secure table mobility, set caster lock pedal to LOCK
position.
AC Power:
1) Plug power cord into properly grounded outlet.
2) Protect cord from damage by casters or other
equipment.
Controls:
1) Emergency Stop Pushbutton disables all tabletop motion
control when engaged.
2) If tabletop motion controls fail to operate, make sure
that the Emergency Stop Pushbutton is not pushed in.
BATTERY OPERATION (DC POWER):
1) The table is FULLY functional, but motions are slower.
2) Connect to AC power for a minimum of 8 hours daily to
ensure a proper battery charge.
3) Failure to push in Emergency Stop Pushbutton when
table is not in use and not connected to AC power for
longer than 72 hours, may result in permanent damage
to the batteries.
CPR:
Return table to level, retract tabletop to minimize
overhang and lower table position to a
comfortable height to perform CPR.
L100-2859A
03/2018
manual prior to use
BRAKES:
The ISR G3 table braking system utilizes a passive magnetic brake;
it remains engaged in the event of a complete loss of electrical
power. Power is necessary to release brakes.
Braking force is limited by the magnet strength. Use caution when
applying pressure to the side of the table as the brakes may be
overcome if more than 70 pounds of force is applied.
To reduce potential motion during patient transfers with larger
patients, release the table brake and move the top in the direction
the patient will be pulled during loading or unloading until the table
top transverse travel limit is reached. Release the panning handle to
re-engage the lock and then transfer the paitent.
PEDESTAL END OF TABLETOP
EMERGENCY
STOP
PUSHBUTTON
POWER
CORD
RECEPTACLE
CASTER LOCK PEDAL
CLEANING AND STORAGE:
1) Unplug power cord and wrap it around cord holder.
2) Push in Emergency Stop Pushbutton.
3) Set caster lock pedal to LOCK position.
4) Clean and disinfect table.
5)
www.imagediagnostics.com
IMAGING END OF TABLETOP
CONNECTION POINT FOR
HAND OR FOOT CONTROLS
CONNECT HERE

Advertisement

loading

Summary of Contents for idi Aspect ISR G3 Q-CARD

  • Page 1 Aspect ISR G3 Q-CARD www.imagediagnostics.com BRAKES: WARNING! Read operator The ISR G3 table braking system utilizes a passive magnetic brake; manual prior to use it remains engaged in the event of a complete loss of electrical power. Power is necessary to release brakes.
  • Page 2 Aspect ISR G3 Q-CARD www.imagediagnostics.com CASTER LOCK TABLE TOP MOTION Table Top Motion PEDAL POSITIONS PANHANDLE CONTROL PANHANDLE SWITCH Handset Control KNOB/ TABLETOP LOCKOUT (Optional) FLOAT BRAKE PUSHBUTTON (Lift underside of Lower FLOAT Elevate Panhandle Knob to (Amber light (Movable in float table top.