- The device is capable of delivering current values exceeding 10mA.
- Before each use always check the integrity of the device. This is a
fundamental requirement for carrying out the therapy; do not use the
device if the buttons or the cables are defective or malfunctioning.
It should be used only by people over 18 and, in any event, who are able
to understand and take action.
- It should not be used for purposes other than transcutaneous
neurostimulation.
- It must be used following the indications and under the strict control of
the physician or qualified physiotherapist.
- The device must be used with the transcutaneous neurostimulation
electrodes suitable for this use.
- The device must be kept out of the reach of children.
- With its current, it can disturb ECG monitoring devices.
- It must not be used in a transthoracic mode as it could cause cardiac
arrhythmia by superimposing its frequency to that of the heart. (Do not
perform the treatment on the chest and the back simultaneously).
- If there is any health problem, it must be used only after consulting a
physician.
- A simultaneous connection of a patient to a high frequency
electrosurgical device can cause burns near the electrodes of the
stimulator and therefore the stimulator may be damaged.
- Once you have turned the device on, make sure the display shows
the software version and the device model: it means that the device is
working and ready to be used;
If it does not, or the display does not show all the segments, turn it off and
on again. If the problem persists, contact our help center and do not use
the device.
- The sudden shutdown shortly after the starting indicates a low battery
level. Recharge as reported in the section "HOW TO CHARGE THE
BATTERIES".
Warnings during the use
While using the electrostimulator, some indications should be followed:
- In the case of damaged cables, they must be replaced with original
parts and not used anymore.
- Only use Globus marked electrodes.
- Great attention has to be paid when current densities exceed 2mA/cm²
(effective value) for each electrode.
The device must be kept out of the reach of pets, as they could damage
the device and contaminate the electrodes and other accessories with
parasites.
45