During running of the battery emulation file, press [Shift]+[I-set](Function) com-
posite key, and the instrument interface displays the running state control inter-
face of the existing file. The user can press Stop to stop existing test and turn
off the battery emulation function. Or, press Reset to restart a new test. When
Reset is selected, the user needs to reset the state of charge of the existing bat-
tery and re-start a new test.
During the charge and discharge test, battery overcharge and overdi-
scharge protection is supported. When the charging capacity is higher
than 110% or the discharging capacity is lower than –10%, the test would
be stopped automatically.