6.7.3 Grad cursor measurement (on B-H graph)

① If
is rotated when Grad cursor is displayed, Grad cursor rotates and the incline value of the cursor (i.e. ratio permeability:
B/( 
・H)) is displayed. (B, H: Magnetic flux density and magnetic field strength at the center of cursor rotation, μ
0
② The rotation center of Grad cursor is the center of Cross cursor. Therefore, if the position of the rotation center of Grad cursor is to be
changed, change the cursor to Cross cursor and move the cursor position.
➂ Pushing
allows the movement amount of the cursor to toggle between Coarse (movement in unit of 20 points) and
Fine (movement in unit of 1 point). For Fine movement, "Fine" is displayed at the upper side on B-H graph.
Fine not displayed (Coarse)
Cursor measurement value
Grad cursor
Fig.6-8 Grad cursor measurement on B-H graph
39
B-H ANALYZER SY-8218 / SY-8219
: vacuum permeability)
0