4.5 Connection of equipments

● The following shows how to connect equipments necessary for measurement.
Connecting the main unit and the measurement POD.
The following is the procedure to connect the measurement POD to the main unit.
① Insert the measurement POD into the opening on the front of the main unit and press it into the bottom.
② Fix the measurement POD by clockwise rotating two clamp knobs on the measurement POD simultaneously. (See 1.3 Cautions for
transportation" and Fig.1-1.)
When installing, removing, or replacing POD, make sure to turn off POWER switch on the front panel of this product; i.e. nothing
appears on LCD screen.
Otherwise, this product may fail.
Connecting the power amplifier
The following is the procedure to use a combination of the power amplifier and this product.
① Connect OSC OUTPUT at the lower right
on the front panel of the main unit with
INPUT of the power amplifier
using the provided OSC cable (BNC-SMA).
Connect them with an optional OSC cable for
SY-911 and IE-1125 when the power amplifier
is IE-1125 or IE-1125x.
② Connect OUTPUT of the power amplifier
with POWER INPUT at the left on the
measurement POD using the provided
power amplifier cable (BNC-BNC).
If a power amplifier is not used, connect
OSC OUTPUT at the lower right on the front
of the main unit with POWER INPUT at the left
of the measurement POD using the provided
OSC cable (BNC-SMA).
Before making connection, make sure to read the instruction manual of the power amplifier to be used. If reverse connection is
made for input/ output of the power amplifier or this product, fatal damage to this product or the power amplifier may be caused.
B-H ANALYZER
OSC OUTPUT
POWER
INPUT
POWER AMP CABLE
(BNC-BNC)
Fig.4-7 Connection of equipments