TEL: +(90) 312 399 28 30 (pbx) FAX: +(90) 312 399 21 97 E-mail: sales@nuve.com.tr WARRANTY CERTIFICATE Nüve warrants that the equipment delivered is free from defects during material and workmanship. This warranty is provided for a period of two years. The warranty period begins from the delivery date.
Page 3
Information contained in this document is the property of Nüve. It may not be duplicated or distributed without its permission. Lifetime of the device is 10 years. PLEASE REGISTER ONLINE TO VALIDATE WARRANTY: To register your warranty online, please visit our web page www.nuve.com.tr and fill in WARRANTY REGISTRATION FORM.
SECTION 1 INTRODUCTION 1.1 USE AND FUNCTION The EN 300, EN 400, EN 400 P, EN 500 and EN 500 P incubators are designed to incubate samples in biological, medical and pharmaceutical laboratories and in many industrial control laboratories. They maintain drying and incubation temperatures between 5°C above the ambient temperature and 80°C and keep the temperature stable within the given tolerances.
K 23 047 Shelf carrier for EN 400 K 23 048 Shelf carrier for EN 400P K 23 046 Shelf carrier for EN 500 K 23 040 Shelf carrier for EN 500P NOTE: 2 pcs. shelf carrier should be ordered for each shelf. 2.3 GENERAL PRESENTATION Display and control panel Door...
SECTION 3 INSTALLATION PROCEDURE 3.1 LIFTING AND TRANSPORT Because of the heavy weight of the incubator, all lifting and transport must be carried out using proper handling equipment. The incubator must be supported from underneath and never turned over. 3.2 UNPACKING Remove the packing cardboard box and the second nylon packing around the incubator.
3.4 MAINS SUPPLY The incubator requires 230 V, 50 Hz. Please make sure that the supplied mains matches the required power ratings. Always plug-in the ıncubator to correctly grounded sockets. A supply fitted with a circuit breaker should be used for protection against indirect contact in case of an isolation fault.
4.2 DISPLAY AND CONTROL PANEL 6. “Operating the program” lamp 1. Temperature display 11. Back / Alarm mute button 2. Time display 7. Data transfer lamp 3. Encoder button 8. Alarm led 9. “End of the program” lamp 4. Start / Stop button 5.
05-Heating Lamp The led is “on” during heating, it indicates that the heating process is carried out. 06-“Operating The Program” Lamp A lamp indicating that the program is running as soon as the device is started. 07-Data Transfer Lamp The lamp indicates that the records are transferred in the memory or the file is transferred during software update.
Again press the encoder button and confirm the set value. Press the back button to return the work screen. 4.3.1 oP Operator Menu Parameters 1: Recording Period: This time is recording period of temperature and error information. 2: Lid Alarm Time: Not used for all EN models. 3: Lid Alarm Range: Not used for all EN models.
Page 13
See the parameter flashing on the temperature display. By turning the encoder button set operating temperature value. Push the encoder button and save set value. See the parameter flashing on the temperature display. By turning the encoder button set operating Set alarm value.
By turning the encoder button set operating time value (01 minute to 99 hours 54 minutes or Hold). Push the encoder button and save set value. See 'dLY' in the temperature display. By turning the encoder button set operating delay time value. If 'Off' is selected, heating will start without delay.
The incubator does not require any periodical maintenance which is carried out by the operator. Please contact to Nuve agent for an authorised service or maintenance 5.2 CLEANING After unplugging the equipment and the equipment is at the room temperature, wipe down the incubator chamber to remove any undesirable effects of the operation, for example spillage.
This error code appears if any probable power cut. “EoF” flashes and the audible alarm sounds on the temperature display. IN CASE OF ANY ERROR, THE PROGRAM İS STOPPED AUTOMATİCALLY AND İMMEDİATELY. PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICAL HELP IF AN ERROR OCCURS.