Tecfluid FLOMAT FX FlomatFX Instruction Manual page 8

Electromagnetic insertion sensor
Table of Contents

Advertisement

Sensor
Fitting
DN
length
length
80
100
125
150
200
101
250
300
350
400
500
600
700
206
800
900
1000
1200
1400
1600
356
1800
2000
Example:
Suppose a pipe of 300 mm inner diameter (DN300) and 5.5 mm thick. In the table we can
see that the distance that the insertion fitting must extend above the exterior wall is H =
60,5 - s = 60,5 -5,5 = 55 mm.
The values in the table are calculated with the gasket supplied with the instrument, which
is 3 mm thick. If the thickness of the gasket is changed, the value of H will change.
The equation to calculate H' for a gasket of thickness d is the following:
In the previous example, if the gasket was 5 mm thick, the distance that the fitting should
extend above the exterior wall would be H' = 55 + 3 - 5 = 53 mm.
8
C
(H)
10
88-s
12,5
85,5-s
15,5
82,5-s
19
79-s
25
73-s
93
31
67-s
37,5
60,5-s
44
54-s
50
48-s
62,5
140,5-s
75
128-s
87,5
115,5-s
145
100
103-s
112,5
90,5-s
125
78-s
150
203-s
175
178-s
200
153-s
190
225
128-s
250
103-s
Qnom
3
m
/h
90,5
141,3
220,9
318,1
565,5
883,6
1.272.,3
1.731,8
2.261,9
3.534,3
5.089,4
6.927,2
9.047,8
11.451,1
14.137,1
20.357,5
27.708,8
36.191,1
45.804,4
56.548,7
H' = H + 3 - d