Abbott Merlin 2 PCS User Manual page 30

2 patient care system
Table of Contents

Advertisement

Table 5. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF
and radiated RF)
Test
IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.7 GHz
*Electromagnetic site survey- Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
9
structures, objects, and people.
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
10
Immunity
7
Compliance Level
(3 Vrms)
[V1=3]
3 V/m
[E1=3]
26
Electromagnetic Environment -
Guidance
8
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Merlin™ 2 PCS, RF Antenna, and
Merlin PSA, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey*, should
be less than the compliance level in
each frequency range.** Interference
may occur in the vicinity of equipment
marked with the following symbol:
9
10