Merlin™ Psa Model Ex3100 Setup; Rf Communication Setup - Abbott Merlin 2 PCS User Manual

2 patient care system
Table of Contents

Advertisement

2.
Select the Interrogate Monitors button.
3.
The programmer will look for advertising pulses from the Confirm Rx™ Insertable Cardiac
Monitor.
Merlin™ PSA Model EX3100 Setup
Connect the Merlin™ PSA Model EX3100 into the PSA port on the Merlin 2 PCS
For more information on the Merlin PSA Model EX3100, including connection of the patient
cables and patient-cable adapters, refer to the on-screen help on the Merlin 2 PCS and to the
Merlin PSA user's manual.
Figure 7. Merlin™ PSA Model EX3100
1. Functional Status LEDs
2. Pace and Sense LEDs
3. Receptacle for patient cable or patient-cable adapter
4. Connector to Merlin 2 PCS

RF Communication Setup

Some St. Jude Medical implantable devices can communicate with the Merlin™ 2 PCS via radio
frequency (RF), which allows for a greater distance between the device and the Merlin 2 PCS. To
establish RF communication, follow the instructions below.
1.
Connect the Merlin™ Antenna to the RF Telemetry port on the back of the Merlin 2 PCS
(see figure below).
NOTE: Connect the Merlin Antenna only to the RF Telemetry port on the back of the
programmer. Do not connect the Merlin Antenna to any other ports.