Abbott Merlin 2 PCS User Manual page 29

2 patient care system
Table of Contents

Advertisement

Table 4. Guidance and manufacturer's declaration — electromagnetic immunity
Test
Electrical fast
transient/burst IEC
5
61000-4-4
Surge
IEC 61000-4-5
Power Frequency
(50/60 Hz) Magnetic
Field, IEC 61000-4-8
Voltage dips, shorts,
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Table 5. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF
and radiated RF)
Test
IEC 60601
Test Level
Applies only to the Merlin 2 PCS.
5
UT is the a.c. mains voltage level prior to application of the test level.
6
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
7
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
8
structures, objects, and people.
IEC 60601 Test Level (Actual
Level)
4
±1 kV for input/output lines
(±1 kV for input/output lines)
±2 kV for power supply lines
(±2 kV for power supply lines)
±1 kV line[s] to line[s]
(±1 kV line[s] to line[s])
±2 kV line[s] to earth
(±2 kV line[s] to earth)
30 A/m
(30 A/m)
6
<5% U
[>95% dip in U
T
0.5 cycle
(<5% U
[>95% dip in U
T
0.5 cycle)
40% U
[60% dip in U
T
5 cycles
(40% U
[60% dip in U
T
5 cycles)
70% U
[30% dip in U
T
25 cycles
(70% U
[30% dip in U
T
25 cycles)
<5% U
[>95% dip in U
T
(<5% U
[>95% dip in U
T
5 s)
Immunity
7
Compliance Level
Electromagnetic Environment —
Guidance
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
] for
Mains power quality should be that
T
of a typical commercial or hospital
] for
environment.
T
If the user of the Merlin™ 2 PCS
requires continued operation during
power mains interruptions, it is
] for
T
recommended that the
Merlin 2 PCS be powered from an
] for
T
uninterruptible power supply or a
battery.
] for
T
] for
T
] for 5 s
T
] for
T
Electromagnetic Environment -
8
Guidance
25

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mer3700

Table of Contents