Measurement Guidelines

Use these guidelines to position the jaws:
• Center the conductor inside the jaws.
• Position the Probe perpendicular to the conductor.
• Make sure that the arrow marked
points toward the correct direction.
Use these guidelines to take measurements:
• If possible, avoid measurements close to other current-carrying
conductors.
• On the Probe, the 100 mV/A range offers the best accuracy.

Maintenance

Before each use, inspect the Probe. Look for cracks or missing portions of the
housing and output cable insulating cover and for loose or weakened
components. Pay particular attention to the insulation surrounding the jaws. If
a Probe fails this inspection, tape it shut to prevent unintended operation. To
check Probe performance, complete the Performance Verification procedure.
To avoid electric shock:
• Do not perform any service procedures unless you are
qualified to do so.
• Read the "Safety" information at the beginning of this
instruction sheet before proceeding.
Figure 4. Measurement Setup
Warning
on the jaw of the Probe