Figure A-2 Bidirectional Scan Pattern - Olympus FlexoFORM User Manual

Table of Contents

Advertisement

10-003591-01EN [Q7780031], Rev. 4, January 2020
To set up for a unidirectional scan
 Set the parameters as outlined in "To set the parameters for a symmetrical scan"
on page 25. The setup and functionality are the same.
Bidirectional scan
The advantage of a bidirectional scan is that it is faster when inspecting long parts
(see Figure A-2 on page 76).
To set up for a bidirectional scan
1.
Set the scan encoder and input parameters according to steps 1, 2, and 4 in "To set
the parameters for a symmetrical scan" on page 25.
2.
Set the indexing button's encoder number (2) and type (Clicker):
a) Select Scan > Encoder > Encoder = 2.
b) Select Scan > Encoder > Type = Clicker.
A bidirectional scan differs from a symmetrical scan as follows:
The scan start value can be set to 0 instead of a negative value (Scan > Area > Scan
Start).
The scan position value is not reset to zero when the indexing button is pressed.
76
Appendix A
Scan axis
Index axis

Figure A-2 Bidirectional scan pattern