Service Policy

1175-0013-000
8. Drain the vaporizer as detailed previously in Instruction 2 and measure the
amount of liquid drained off. The amount of liquid consumed should be as
follows:
Enfluratec
Fluotec
Isotec
Sevotec
Appropriate action must be taken to handle the exhaust gases and spillage.
Note
This method is designed to be a quick and easy check of vaporizer operation and,
therefore , it is somewhat imprecise. However, it is unusual for the measured liquid
consumption to vary by more than 25% of the values listed above.
Repairs and service procedures must be performed at a Datex-Ohmeda Authorized
Service Center. Contact your Datex-Ohmeda Service Representative or Datex-Ohmeda
Authorized Distributor for information on maintenance and shipping.
15.5 milliliters
13.5 milliliters
15.5 milliliters
16.6 milliliters